Section 13 Troubleshooting

Display
Detector overload
cannot read the display
Chemiluminescence line
develops in a bell shaped
curve.
Negative percent inhibition
Database is full. No new
Error
measurements can be saved.
01
System RAM test failed
02
System EPROM test failed
03
LCD display test failed
04
Non-volatile RAM test failed
05
Storage measurements corrupt
06
Configuration settings corrupt
07
Usage counter corrupt
08
A/D input failure
09
Reference LED failure
11
Reference LED reading fault
100
Cannot clear measurements
101
Cannot store range settings
Troubleshooting.fm
Table 4 Luminometer troubleshooting
Fault
Change the measurement range to the 0–2000
range (refer to
range on page
Change batteries (refer to
Luminometer battery replacement on page
The chemiluminescence reagents are weak and
need to be replaced.
Do the reference measurement again.
Salt could be in the sample or the reagents may be
faulty.
Change the contrast.
Erase all measurements (refer to
Erase the results saved on the luminometer on
page
Memory failure. Contact technical support.
Memory failure. Contact technical support.
Continue operation. If fault occurs again, contact
technical support.
Memory failure. Contact technical support.
All saved measurements will be erased. Continue
operation. If fault occurs again, contact technical
support.
LCD contrast and measurement range will be
reset. Continue operation. If fault error occurs
again, contact technical support.
Usage counter will be reset to zero. Continue
operation. If fault occurs again, contact technical
support.
Failed measurement. Contact technical support.
Failed measurement. Contact technical support.
Failed measurement. Contact technical support.
Remove power and then apply power to the
luminometer. If fault occurs again, contact
technical support.
Remove power and then apply power to the
luminometer. If fault occurs again, contact
technical support.
Corrective action
section 3.2.3, Set the measurement
18).
17).
section 12.3.1,
112).
section 3.2.2,