Specifications are subject to change without notice.
Specification | Details |
Dimensions (W x D x H) | 37 x 30 x 170 mm (1.45 x 1.18 x 6.69 in.) |
Enclosure rating | IP67 |
Weight | 135 g (0.297 lb) with batteries |
Power requirements (internal) | AAA alkaline batteries (4x) |
Battery life | Pocket Pro: 450 hours Pocket Pro+: 450 hours (200 hours with backlight on) |
Operating temperature | 0 to 50°C (32 to 122°F) |
Storage temperature | –20 to 60°C (–4 to 140°F) |
Operating humidity | 80% (non-condensing) |
Altitude | 2000 m (6562 ft) |
Instrument range | 0.0 to 14.0 pH |
Resolution | Pocket Pro: 0.1 pH; Pocket Pro+: 0.01 pH |
Accuracy[1] | Pocket Pro: ±0.1 pH; Pocket Pro+: ±0.02 pH |
Calibration points | Pocket Pro: 3 points (auto); Pocket Pro+: 3 points (auto) or 2 points (custom) Note: Temperature calibration is not available. |
Auto-recognition calibration standards | USA: 4.01, 7.00, 10.01 pH; NIST: 4.01, 6.86, 9.18 pH |
Certifications | CE mark, FCC, Industry Canada, KC Mark, RCM, China RoHS |
Warranty | 6 months for Pocket Pro tester, 1 year for Pocket Pro+ tester and 6 months for replacement sensor for manufacturing faults only. Damage from use is not covered. |
1 Based on a 3-point calibration and the calibration standards at the same temperature as the samples measured. Also valid for 5.5 to 8.5 pH based on a 1-point calibration, 0.0 to 8.5 pH based on a 2-point calibration with pH 7 and pH 4 standards, or 5.5 to 14 pH based on a 2-point calibration with pH 7 and pH 10 standards.
In no event will the manufacturer be liable for damages resulting from any improper use of product or failure to comply with the instructions in the manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer's website.
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information. | |
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user. |
Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class B:
Supporting test records reside with the manufacturer.
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
The Pocket Pro™ pH tester and Pocket Pro™+ pH tester measure the pH of general water samples. These testers are waterproof and float.
|
|
|
|
|
|
|
|
Make sure that all components have been received. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
|
|
|
Explosion hazard. Incorrect battery installation can cause the release of explosive gases. Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation. Do not mix new and used batteries.
Fire hazard. Battery substitution is not permitted. Use only alkaline batteries.
Install the four AAA alkaline batteries in the tester. Refer to the illustrated steps in Figure below.
Figure below shows the measurements, calibration standard information and indicator icons shown on the display.
Table 1 gives descriptions of the icons.
|
|
|
|
|
|
|
| |
|
|
2 Shows after auto calibration
3 Shows during auto calibration
4 Shows during custom calibration.
Table 1 Display icons
Icon | Description |
Shows the battery power level. Flashes when the battery power is less than 10%. | |
Shows when the lock feature is on. When the lock feature is on, the parameter value on the display does not change. Push to set the lock feature to on or off. | |
Sensor? | Refer to Troubleshooting |
Shows when the tester is in calibration mode or when a calibration is due. If "?" shows next to the calibration icon, the last calibration was not successful. When the ACAL setting is set to Yes (default), "?" shows next to the calibration icon when a calibration is due. |
Figure 5 and Figure 6 show the Pocket Pro and Pocket Pro+ keypads. Table 2 gives the key descriptions.
Figure 5 Keypad overview - Pocket Pro
|
|
|
Figure 6 Keypad overview - Pocket Pro+
|
|
|
Table 2 Key functions
Key | Description |
Push and hold to set the power to on or off. | |
Push and hold to set the power to on or off. Push to set the backlight to on or off. After 1 minute of no activity, the backlight switches off. | |
Push to start a calibration. To exit a calibration, push and hold. Push and hold until "SEt" shows on the display to go to the settings menu. To exit the settings menu, push and hold until "End" shows on the display. When in the settings menu, push to scroll through the settings. Note: Power cannot be set to off while in settings or calibration mode. | |
Push to set the lock feature to on or off. When the lock feature is on, the lock icon shows and the parameter value on the display does not change. |
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
Calibrate the tester before initial use and when:
Calibrate the tester with the:
Before the initial calibration and after a dry storage, soak the sensor for several minutes in the sample or tap water.
5 Custom calibration can only be done on the Pocket Pro+ tester.
Items to collect: One, two or three auto-recognition calibration standards
(Pocket Pro+ only)
Items to collect: One or two calibration standards or samples of known pH value
Note: Air bubbles under the probe tip when submerged can cause slow stabilization or error in measurement. Shake the tester from side to side to remove air bubbles.
Option | Description |
Unit | Select the temperature unit that shows on the display— C (Celsius) or F (Fahrenheit). |
bUFr | Select the pH buffers that are used for auto calibration—USA (4.01, 7.01, 10.01, default), NIST (4.01, 6.86, 9.18) or Cus (custom standard). Note: "Cus" can only be selected on the Pocket Pro+ tester. |
AOFF | Set the auto-off feature to On (enable, default) or Off (disabled). When set to On, power is automatically set to off after 8 minutes of no activity. |
ACAL | Enable or disable pH calibration reminder—Yes (enable, default) or No (disable). When set to Yes, "?" shows next to the calibration icon on the pH screen when a pH calibration is due. |
rSEt | Change the settings to the factory defaults—Yes or No (default). When set to Yes, changes the settings to the factory settings and default values. |
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Clean the sensor when:
Table 3 Cleaning agents
Contaminant | Cleaning agent | Time |
Grease, oils and fats | Electrode cleaning solution | 2 hours maximum |
Mineral buildup | 10% hydrochloric acid (HCl) solution | 5 minutes maximum |
When the battery icon flashes or the tester will not come on, replace all four batteries. Refer to Install the batteries.
Note: Only Pocket Pro+ testers have replaceable sensors.
To replace the sensor, refer to the instructions supplied with the sensor.
Message | Possible cause | Solution |
SENSOR? | The calibration slope is ±10–15%. | Gently clean the sensor. Refer to Clean the sensor. |
ECAL | Calibration failure. The pH calibration slope is greater than ±15%. | Gently clean the sensor. Refer to Clean the sensor. Calibrate again. If calibration failure continues, replace the tester or sensor as applicable. |
"- - - -" (top line) | The parameter value is not within the measurement range of the tester. Refer to Specifications. | Make sure that the value of the sample is within the measurement range of the tester. Gently clean the sensor. Refer to Clean the sensor. Calibrate as necessary. |
"- - - -" (bottom line) | The temperature value is not within the operating temperature range of the tester or a temperature sensor failure occurred. Refer to Specifications for the operating temperature range. | Make sure that the sample temperature is within the operating temperature range of the tester. Contact technical support as necessary. Note: The tester can still be used if a temperature sensor failure has occurred, but without automatic temperature compensation. |
Battery icon flashes | The batteries have less than 10% power remaining. | Replace all four batteries. Refer to Install the batteries. |
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Table 4 Replacement parts
Description | Quantity | Item no. |
AAA alkaline batteries | 4/pkg | 4674300 |
Lanyard | 1 | 201305 |
SINGLET, 7.00 pH | 20/pkg | 2770120 |
pH sensor, replacement | 1 | 9532001 |
Table 5 Accessories
Description | Quantity | Item no. |
SINGLET, 4.01 pH | 20/pkg | 2770020 |
SINGLET, 10.01 pH | 20/pkg | 2770220 |
Electrode cleaning solution | 500 mL | 2965249 |
Hydrochloric Acid (HCl) | 2.5 L | 13406 |
Hydrochloric Acid (HCl) | 500 mL | 13449 |
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. (970) 669-3050
(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
orders@hach.com
www.hach.com
HACH LANGE GMBH
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
info-de@hach.com
www.de.hach.com
© Hach Company/Hach Lange GmbH, 2013–2014, 2023. All rights reserved.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the Pocket Pro that isn't answered in the manual? Leave your question here.