3.2 Prepare the luminometer for use

Do the procedures in this section before each deployment to prepare the
luminometer for use.
3.2.1

Test the operation

Do this procedure to make sure that the luminometer is operating correctly. If the
luminometer passes all the tests done in this procedure, it is operating correctly.
To test the operation of the luminometer:
1.
Open the hinged cell lid of the luminometer and make sure that a sample is not
in the cell.
2.
Remove the black test tube holder from the cell. Make sure that the cell is clean
and free from debris.
3.
Put the black test tube holder into the cell and close the cell lid.
4.
Push ON (green button) for several seconds to apply power to the instrument. If
the instrument does not energize, replace the batteries (refer to
page
112).
16
Figure 2 Function keys
Function
Does the action on the display directly above the key.
Does the action on the display directly above the key.
Illuminates the display.
Removes power to the instrument.
Applies power to the instrument.