1. Using a calibrated pressure gauge, check the supply cylinder to ensure that
it contains 3000 PSI (207 BAR). It is very important to ensure that the EBIS is
filled to its total capacity but not overfilled.
2. Remove the protector cap from the threaded male fitting of the fill adapter.
Inspect the fill adapter to ensure that the o-ring is present and seated evenly at
the base of the threads. Check the threads of the male fitting, making sure they
are clean and not damaged.
3. Screw the threaded male fitting on
the fill adapter into the fill port. DO
NOT apply a wrench or otherwise
overtighten the fill adapter (Fig. 7).
4. Loosen the fill adapter yoke screw
as needed so that the dust cap can be
removed from the inlet fitting.
5. Inspect the supply cylinder o-ring
to make sure it is present, in good
condition and seated evenly in the
supply cylinder valve.
6. While supporting the EBIS with one
hand, place the yoke of the fill adapter
over the cylinder valve to align the inlet
fitting flush against the valve o-ring.
Screw the fill adapter yoke screw into
the small dimple on the backside of
the cylinder valve until finger-tight
(Fig. 8).
7. Close the bleed valve screw on the
fill adapter (Fig. 8).
EBIS - 3000 & 4500 User's Manual
Figure 7
Bleed Valve Screw
Figure 8