4.5 Emergency Stop

IMPORTANT
When the emergency stop is activated, the CPR
function can still be activated, by pressing button
A, to lower the wings to horizontal position if
required.
When the emergency stop is activated it stops all controls
and movements immediately (wings will stay in current
position).
1. To activate, push the button B.
2. To deactivate, rotate the button B clockwise.
3. Press button A to reset the system.

4.6 Manual operation

WARNING!
Risk of squeezing and injury
If both wings are simultaneously elevated to
angles of more than 12° there is a risk of
squeezing the patient.
– This option must only be used by trained
personnel.
The first time MANUAL is activated, after the system
has been switched on, it will reset itself before the
function can be used.
1. Press and hold button H for 3 sec. to activate manual
operation.
2. Press buttons on hand control as required.
To raise the left wing, press I
To lower the left wing, press J
To raise the right wing, press K
To lower the right wing, press L
-Left and right buttons can be activated at the
same time to move both wings simultaneously.
-The wings can be moved in a range from 0° to
60°.
3. To deactivate manual operation, press button H or the
ON/OFF & CPR button A.
Status indicators
1605513-C
Display
0 = no activity
U = wing moves up
d = wing moves down

4.6.1 Recommended Procedures

Transfer from flat position to the side:
1. Raise both wings simultaneously to increase the comfort
for the end user.
2. When both wings are slightly raised, continue raising
only the primary wing and leave the secondary wing
in position.
3. When the primary wing has reached the desired height,
you can (but don't have to) lower the secondary wing
to horizontal.
Transfer from one side to the other:
1. Simultaneously raise the lower side and lower the higher
side
2. Stop lowering the secondary wing when it is still slightly
raised and continue raising only the primary wing.
3. When the primary wing has reached the desired height,
you can (but don't have to) lower the secondary wing
to horizontal.

4.7 Automatic operation

The AUTO program is pre-defined and can not be changed.
Program 1 and 2 can be individually defined by the care staff
(see chapter Programming). Upon delivery the programs are
empty and need to be defined before usage.
-If one of the programs is activated, the system will
reset to FLAT position before starting the selected
program.
-If a new program is selected while another program
is active, the system will reset to FLAT position before
starting the new program.
1. Press program buttons as required.
To start the AUTO Program, press and hold M for
3 sec.
To start Program 1, press and hold N for 3 sec.
To start Program 2, press and hold O for 3 sec.
2. To stop a running program press the ON/OFF & CPR
button A or press button M N or O, depending on
which program is currently active.
The hand control will not work when a program is
active.
LED
MANUAL LED active
Usage