When working with RNA samples, make sure to consider following
recommendations:
• Always follow the GLP-rules established in the laboratory
• Wear gloves all the times to prevent contamination with RNases
• Thaw RNA samples and ladder on ice and keep them on ice during sample
• Ensure all working areas and plastic ware are RNase free
Not following these general guidelines might result in RNA degradation and false data.
N O T E
Figure 26
100
preparation
Example of a degraded RNA sample in the High Sensitivity RNA assay
resulting in missing peak identification in the Gel View (Left) and
Electropherogram View (Right) of the TapeStation Analysis software.