Keysight M8194A User Manual page 141

120 gsa/s arbitrary waveform generator
Table of Contents

Advertisement

Parameter Suffix
Parameter Suffix
Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide
:SYSTem:ERRor[:NEXT]?
Query
:SYSTem:ERRor[:NEXT]?
Parameters
None
None
Description
Read and clear one error from the instrument's error queue.
A record of up to 30 command syntax or hardware errors can be stored in
the error queue. Errors are retrieved in first-in-first-out (FIFO) order. The
first error returned is the first error that was stored. Errors are cleared as
you read them.
If more than 30 errors have occurred, the last error stored in the queue
(the most recent error) is replaced with "Queue overflow". No additional
errors are stored until you remove errors from the queue.
If no errors have occurred when you read the error queue, the instrument
responds with 0,"No error".
The error queue is cleared by the *CLS command, when the power is
cycled, or when the firmware is re-started.
The error queue is not cleared by a reset (*RST) command.
The error messages have the following format (the error string may contain
up to 255 characters):
error number,"Description", e.g.
-113,"Undefined header".
Examples
Query
:SYST:ERR?
:SYSTem:HELP:HEADers?
Query
:SYSTem:HELP:HEADers?
Parameters
None
None
Description
The HEADers? query returns all SCPI commands and queries and IEEE
488.2 common commands and common queries implemented by the
instrument. The response is a <DEFINITE LENGTH ARBITRARY BLOCK
RESPONSE DATA> element. The full path for every command and query is
returned separated by line feeds. The syntax of the response is defined as:
Remote Programming
141
4

Advertisement

Table of Contents
loading

Table of Contents