Ph Calibration - Hach TitraLab User Manual

Ph & acid content analyzer
Hide thumbs Also See for TitraLab:
Table of Contents

Advertisement

4.6

pH calibration

1. Place a magnetic stirring bar and 10 mL of the pH 7.00 calibration solution in a
titration cell.
2. Attach the titration cell to the magnetic stirrer.
3. Lower the electrode into the titration cell.
Hinweis: Make sure that the electrode cannot be damaged by the stirring bar or come into
contact with the bottom of the titration cell. Make sure the combined pH Electrode, including the
reference element porous pin is covered by sample.
4. Select
PH CALIBRATION
shown on the display.
5. Select
.
START
6. Following successful calibration, "
the display.
FINISH CALIBRATION
: Place a magnetic stirring bar and 10 mL of the pH 4.00 calibration
NEXT POINT
solution in a titration cell. Continue as described.
7.
"IMMERSE THE PH ELECTRODE IN THE SECOND BUFFER."
8. Select
.
START
9. Following successful calibration, "
10. Press
to confirm.
OK
Note: Use the lids supplied to close the titration cells containing the calibration solutions. The
calibration solutions can be used several times.
.
"IMMERSE THE PH ELECTRODE IN THE FIRST BUFFER."
" or "
NEXT POINT
: Calibration finished as one-point calibration.
FINISH CALIBRATION
Operation
" is shown on
FINISH CALIBRATION
is shown on the display.
" is shown on the display.
is
25