Ph Probe; Ph Probe Connection; Ph Probe Programming; Ph Probe Calibration - Hach Sigma 950 Instrument Manual

Flow
Hide thumbs Also See for Sigma 950:
Table of Contents

Advertisement

Optional Device Installation

7.2 pH Probe

7.2.1 pH Probe Connection

7.2.2 pH Probe Programming

7.2.3 pH Probe Calibration

50
The pH probe consists of five wires, three for the pH probe and two for the temperature
probe. Since the pH probe reading needs to compensate for temperature variation, there
is a temperature probe built into every pH probe.
1. Attach the clear wire to either screw on the terminal strip labeled GLASS.
2. Attach the black wire on the shield of the cable to the REF screw on the other
terminal strip.
3. Attach the red wire to the GND screw on the terminal strip.
4. Attach the green and yellow wires to the screws labeled RTD (Resistance
Temperature Detector).
Note: The green and yellow wires can be attached to either one of the RTD terminal screws
because there is no polarity present.
1. From the Main Menu, select
1. Highlight Select Inputs using the
2. Highlight pH using the
3. Press
CHANGE CHOICE
.
ACCEPT
4. Enter a logging interval using the numeric keypad, then press
intervals are shown on the status bar.
5. Select another channel to configure or press
function key to return to the Main Menu.
MAIN MENU
Once the pH probe is connected and programmed, calibrate the pH probe. Calibrating
the pH probe requires a thermometer and any two of the following buffer solutions: 4, 7,
or 10 pH. The pH probe is an application sensitive device. When used in harsh
environments, the accuracy and life expectancy of the probe decreases.
Calibrate the pH probe each time they are cleaned or replaced. Regular inspection and
comparison to a hand-held pH meter can help determine the optimum cleaning and
calibration schedule for your application.
1. From the Main Menu, select
2. Place the pH probe into the first buffer solution. Press any key to continue.
Table 17 pH Connector Pin Assignments
Pin
A
B
C
D
E
F
OPTIONS>ADVANCED OPTIONS>DATALOG.
and
UP
DOWN
and
keys, then press
UP
DOWN
to cycle between Logged and Not Logged, then press
OPTIONS > ADVANCED OPTIONS > CALIBRATION > pH.