PRINCIPLE OF OPERATION:

The UltraMaxO
Oxygen Analyzer measures oxygen concentration and flow using
2
ultrasound technology and measures pressure using a piezoresistive silicon
pressure sensor.

WARNINGS:

Indicates a potentially hazardous situation, if not avoided, could result in death or
serious injury.
Not for use in an MRI environment.
Improper use of the UltraMaxO
improper treatment and/or patient harm. Follow the procedures outlined in this
user manual.
The UltraMaxO
DO NOT use the UltraMaxO
DO NOT use the UltraMaxO
when flowing at rates lower than its optimal performance as specified by the
concentrator manufacturer; generally 4 LPM or less on concentrators that have a
maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a maximum
flow of 5 LPM.
Not for use in anesthesia applications or for measuring oxygen concentration from
any sources other than conventional oxygen concentrators.
Not for use with inhalation agents. Operating the UltraMaxO
explosive environments may result in fire or explosion.
Not suitable for use in the presence of flammable anesthetic mixtures.
Users must become thoroughly familiar with the information contained in this
Operation Manual before use. Strict adherence to the operating instructions is
necessary for safe, effective product performance. This product will perform only as
designed if operated in accordance with the manufacturer's operating instructions.
Use only genuine Maxtec accessories. Failure to do so may seriously impair the
performance of the UltraMaxO
other than an authorized Maxtec service representative could cause the product to
fail to perform as designed.
Use of the UltraMaxO
erratic readings.
If the UltraMaxO
remove the batteries and let the device dry completely. When dry, replace the
batteries and check for proper operation.
DO NOT autoclave or expose the UltraMaxO
DO NOT use ethylene oxide sterilization. performance in flow and pressure readings.
II
!
can cause inaccurate oxygen readings leading to
2
is for checking oxygen concentrators only.
2
for continuous oxygen monitoring.
2
to measure the oxygen concentration of a concentrator
2
. Repair or alteration of the UltraMaxO
2
near devices that generate electrical fields may cause
2
is ever exposed to liquids from spills or immersion, immediately
2
866.4.Maxtec
to high temperatures (>60°C).
2
www.maxtecinc.com
in flammable or
2
by anyone
2