CSA STD C22.2 No. 60601-1 NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.com Product Disposal Instructions: Use only genuine Maxtec accessories and replacement parts. Failure to do so may seri- ◆...
2.0 OPERATING INSTRUCTIONS 2.1 Getting Started 2.1.1 Protect Tape Prior to turning on the unit, a protective film covering the threaded sensor face must be removed. After removing the film, wait approximately 20 minutes for the sensor to reach equilibrium. WWW.MAXTEC.COM • (800) 748-5355...
A simple calibration may be made with the sensor open to static at Ambient air. For optimum accuracy Maxtec recommends that the sensor be placed in a closed loop circuit where gas flow is moving across the sensor in a controlled manner.
Press and hold the Calibration Button for three seconds to manually force a new • calibration. If unit repeats this error more than three times, contact Maxtec Customer Service for possible sensor replacement. CAL ERR HI: Sensor voltage too high Press and hold the Calibration Button for three seconds to manually force a new •...
Clean the sensor with a cloth moistened with isopropyl alcohol (65% alcohol/water ◆ solution). R103P90 Perfusion Tee Adapter Maxtec does not recommend use of spray disinfectants because they can contain salts, ◆ RP16P12 Long-Neck Tee Adapter which can accumulate in the sensor membrane and impair readings.