Instrument Installation
Site Requirements
To install and operate the ANKOM
•
Adequate power (see "Operating Environment" section of this manual)
•
Drain or waste container to capture waste fluids
•
Bench space that can accommodate the instrument dimensions of 132 cm (52") L x 92 cm (35") H x 77 cm (30") D
•
80-100 psi (5.5-6.9 bar) Nitrogen Supply and connection tubing
Instrument Installation Procedure
TDF
To install the ANKOM
Unpack and place the instrument where it will be used.
1.
The instrument shipping container consists of three separate cardboard pieces: a top, a bottom, and a sleeve that forms the
body of the container. Remove the top and the sleeve of the instrument shipping container. Lift the instrument off of the
shipping container bottom and place it on a surface that is firm and level in an area near a Nitrogen Supply and a drain or
chemical waste container.
The instrument must not be subjected to excessive shock, vibration, dirt, moisture, oil, or other fluids.
Unpack the attachments and accessories.
2.
Within the shipping container is a cardboard box containing attachments and accessories. Open the box and verify that all
of the following items are there.
1 – Crescent Wrench (204.9)
1 – 3/8" Nut Driver (8416)
1 - TDF Synthetic Grease (TDF107)
1 – Drain Tray (8002)
Rev 9/20/18
TDF
Dietary Fiber Analyzer you will need the following:
Because the instrument is powered by pneumatics, do NOT run it without an
adequate Nitrogen Supply. You can run approximately 5 to 10 full assays
(IDF/SDF or TDF) with a 304 cu ft tank of nitrogen.
Dietary Fiber Analyzer, follow the steps below.
Waste fluids are drained from the instrument using gravity. Therefore, position the
instrument so that the drain tube can run down from the Waste Tray.
Ethanol is used within this instrument. Follow both local and federal regulations for
vent hood requirements when installing this instrument.
1 – 1/8" Allen Wrench (Z303)
1 – Drain Tube Clamp (6138)
1 – 3/8" wrench (8418)
1 – Drain Hose Assembly (TDF96) that includes these five parts
1 - Adhesive Clamp (8133)
Operator's Manual
1 – T-Handle Torx T20 Driver (Z276)
2 – Spray Tips (TDF44)
1 – Drain Tube (8132)
2 – Drain Tube Elbow Connectors (8288)
pg. 11
Need help?
Do you have a question about the TDF and is the answer not in the manual?