Channel Setup Menu
Channel autoscale
Channel, Bandwidth/Wavelength...., Filter On Off
Wavelength
This function is available on the optical channel only.
The Wavelength function selects the desired wavelength for calibrated mea-
surements. The factory calibrated wavelength is 850 nm. A user-calibrated
wavelength is also available and can be calibrated in the range from 750 nm to
860 nm. Refer to Chapter 3, "Calibration Overview" for additional information
on performing a calibration.
Channel, Bandwidth/Wavelength...., Wavelength
Filter
This function selects the specific filter for the type of compliance testing to be
performed.
Channel, Bandwidth/Wavelength...., Filter, 1063 Mb/s or 1250 Mb/s
Channel autoscale
The Channel autoscale softkey provides a convenient and fast method for deter-
mining the standard vertical scale setting with the highest resolution that will
not clip the waveform. Timebase and trigger settings are not affected.
This function is useful in manufacturing environments where the timebase
and trigger settings remain constant and only the vertical scale needs to be
adjusted for signal level variations in multiple devices under test.
Channel, Channel autoscale
2-8