BeaconMedaes Lifeline Installation, Operation And Maintenance Instructions page 14

Lifeline medical systems. liquid by high pressure fully-automatic manifold
Table of Contents

Advertisement

Liquid by High Pressure Manifold
10.
Each header extension is shipped with a header
bracket. Each header bracket shouldbe attached
to a manifold header stand. Assemble each
header stand by attaching a post base to one
end of 66-inch channel as shown in Figure 8,
Detail A. Use two each 3/8" bolts, flat washers
and channel nuts. Tighten all bolts.
11. Set header stand in a vertical position roughly in
alignment with two vertical control panel stands.
Position header bracket against channel and
on bottom side of header as shown in Figure 5.
Attach bracket to channel as shown in Figure 8,
Detail B. Use two each 3/8" bolts, flat washers
and channel nuts. Tighten all bolts.
12. Align header stand with two control panel
vertical stands, centering the header bracket
between two pigtail connections. Mark locations
of all four base mounting holes. Move header
stand aside and drill holes. Minimum 3/8"
diameter mounting bolts are recommended (by
others).
13. Reposition header stand over holes and install
mounting hardware (by others). Tighten all
mounting bolts.
14. Assemble u-bolt and bracket hardware asshown
in Figure 5. Tighten u-bolt nuts.
15. Additional header extensions may be attached
end to end if required.
16. Install large nut and plug on end of last header
extension. Tighten largenut.
CAUTION:
Do not use thread sealant on header or pigtail
connections.
4107 9013 95.04
Stand Kit, Manifold Control Panel
P/N 6-135000-KT
POST BASE (X2)
80" CHANNEL (X2)
Figure 6
Stand Kit, Manifold Header
P/N 6-135001-KT
POST BASE (X2)
80" CHANNEL (X2)
Figure 7
2-8
3/8-16 CHANNEL NUT (X8)
3/8" FLAT WASHER (X8)
3/8-16 X 1" BOLT (X8)
3/8-16 CHANNEL NUT (X8)
3/8" FLAT WASHER (X8)
3/8-16 X 1" BOLT (X8)