Preparation Of The Measuring System For Po4-P Measurement, Calibration Standards - wtw P 700 IQ Preparation Instructions

Orthophosphate analyzer for measuring phosphate phosphorus
Hide thumbs Also See for P 700 IQ:
Table of Contents

Advertisement

Preparation instructions for P 700 IQ-solutions
2.1
Preparation of the measuring system for PO4-P measurement, calibration
standards
Chemicals required:
• Stock solution: Traceable ion standard PO
(e.g. WTW article no. 250478)
• Fuming hydrochloric acid (HCl) 37 % p.A. (≈ 12 M)
• Demineralized water (H
Aids and appliances:
• Volumetric flask 1000 ml
• Suitable pipettes
• Measuring system for testing for PO4-P (e.g. ion chromatograph Metrohm IC Compact 761)
3-
Production of PO
-P calibration standard:
4
The volumetric flask to be selected as per Table 2.1-1 must be filled to 80 % with
demineralized water. Wait for the temperature adaptation of the water. Add the
corresponding volume of stock solution (see Table 2.1-1) using a pipette. Then pipette the
amount of fuming hydrochloric acid specified in Table 2.1-1. Fill to the mark (demineralized
water) using the spray bottle. Mix the solution thoroughly.
Concentration of the
solution to be tested
3-
in mgl/l PO
-P
4
1.600
8.000
20.00
Table 2.1-1: Volumes of pipette and volumetric flask to be used
If the calibration standards are prepared as described above, calibration standards according
to Table 2.1-2 are obtained using the following concentrations:
Concentration of the
solution to be tested
3-
in mgl/l PO
-P
4
1.600
8.000
20.00
Table 2.1-2: Concentration of the solution to be tested and calibration standard
Shelf life of calibration standard
The volumetric flask must be provided with the measuring ion and its concentration as well
as with the preparation date of the standard and hazardous substance marking and may be
used for one month (storage at room temperature) for calibrating the measuring system.
Preparation of the measuring system for PO4-P testing
Prepare according to operating manual.
3
, 1000 mg/l
4
O); Residual conductivity
2
Volumetric flask
volume to be used in
ml
1000
1000
1000
Concentration of the
calibration standard
3-
in mgl/l PO
-P
4
1.600
8.000
20.000
κ < 2µS/cm
25 °C
Volume of stock
Addition of fuming
solution to be
hydrochloric acid in
pipetted in ml
4.906
24.529
61.323
ml
1.33
1.33
1.33
11