9. Checklist for technical problems

Problem
Shower stretcher
/ dressing table is
hard to move
Shower stretcher
/ dressing table
does not move
upwards or
downwards
Light on the
charger does not
come on
Shower stretcher
dressing table
stops
52 | Instructions for use | Shower stretcher / Dressing table
Cause
Dirt and soap between wheels
Connection between pump
and pedal has become discon-
nected
Pump leaking or broken
Emergency stop button used
Battery empty
Plugs removed
Manual controls not working
Charger is not plugged in
Charger is not working
Emergency stop button used
Technical issue
Solution
Clean wheels
Tighten screws on connector
Contact Lopital Nederland B.V.
Turn emergency button 90
degrees to the right
Charge battery / install reserve
battery
Check plugs (to side of control
box)
Check plugs
Contact Lopital Nederland B.V.
Connect charger
Contact Lopital Nederland B.V.
Turn emergency button 90
degrees to the right
Contact Lopital Nederland B.V.