  • Page 4: Introduction

    1. Introduction 1.1. Introduction The height-adjustable, electric shower and toilet chair Reflex is a medical device for caring for and washing clients in an ergonomically responsible manner . Comfort is the primary consideration for both the client and the carer . The Reflex can also be used as a toilet chair, in combination with a bedpan / bucket .
  • Page 5 • The use of components (chargers, cables, actuators, batteries, etc . ) that have not been supplied by Lopital may result in increased emissions or decreased immunity of the Reflex . • Use the Reflex in a clean and tidy working environment .
  • Page 6: Warranty

    1.3. Warranty Lopital products are guaranteed for 2 years with respect to construction and / or material faults . In the event of varying warranty terms, the exact period will be indicated on your invoice; this also serves as the warranty itself .
  • Page 7 Charger Figure 1 Control panel . Head support complete Handle Back support Armrest (front closure) Frame Seat Filler piece - seat Bedpan guides Tilting motor (Right hand side at rear) Footrest Hinge complete cover (6840 5627) Scissor mechanism Swivel wheel Ø12 .5 cm Bumper Dual braking...
  • Page 8: Activating

    2.2. Activating Before you can use the Reflex, you must activate it by pressing both up arrows for 2 seconds (See figure 1) . 2.3. Working with the Reflex Step 1 Disconnect the Reflex from the charger if necessary . Step 2 Activate the Reflex (See paragraph 2.2 Activating) .
  • Page 9: Operation

    Step 10 Take the chair off the brake when you wish to move it (See paragraph 3.1. Wheel blocking) Always move the chair in the low position. Step 11 When you wish to tilt the chair or move it upwards / downwards, you must apply the brake (See paragraph 3.5 Moving upwards / downwards and paragraph 3.6.
  • Page 10 • The carer or the client first raises one leg; the carer then folds the foot plate down (See figure 5). The same procedure is then carried out with the other leg . Make sure that the footrest is set to the right height . •...
  • Page 11: Armrest With Front Closure

    3.3. Armrest with front closure • If the client has taken a seat on the Reflex, the armrests with front closure can be folded down; these are then turned for the person concerned . • The armrests with front closure are secured in a horizontal position .
  • Page 12: Head / Neck Support

    3.4. Head / neck support • The Reflex has a head / neck support that can be adjusted in terms of both height and depth . • In order to set its height, the rotating button (1) on the back of the chair must be undone .
  • Page 13: Tilting

    3.6. Tilting • The Reflex is equipped with a backwards tilting function of up to 30 degrees . This function is operated by pushing the button that shows an arrow pointing backwards . The chair will then tilt (See figure 15). •...
  • Page 14: Use Of A Bedpan Or Bucket

    • The bedpan or bucket is slid in from the back of the chair (See figure 17) . • If you are using the Lopital plastic bedpan, you must slide it as far forwards as possible (See figure 18) .
  • Page 15 Make sure that you push the plug into the correct position in the charging point. Only the original charger for the Reflex may be used. If you use another charger, Lopital cannot be held Figure 21 Charging point . liable for any consequences, such as damage to the electronics.
  • Page 16: Safety Functions

    4. Safety functions 4.1. Lock button The Reflex is fitted with a Lock button (1) (See figure 23) . Operate the Lock button if the Reflex is not to be used for an extended period . This will prevent any unauthorised use of the chair .
  • Page 17: Cleaning And Maintenance

    5. Cleaning and maintenance The Reflex must be cleaned after every use, with water and regular household cleaning agents . Remove excess soap residues . PUR integrated foam components with IMC coating should be rubbed clean and disinfected . Do not use any abrasive or other aggressive cleaning products (e .g .
  • Page 18 . This process will preferably be carried out by Lopital or another authorised body . Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance .
  • Page 19: Storage And Transport

    (See paragraph 4.1. Lock button). 7. Removal of parts All parts that are replaced or removed can be sent back to Lopital BV . You may also offer the parts to the engineer for destruction . We will ensure that all parts are processed in an environmentally-friendly manner .
  • Page 20: Drive

    8.2. Drive Drive: Actuator 24V Stainless steel spindles for high / low movement 24 V actuator for tilt adjustments Battery: battery capacity 7Ah Charger : 110-240 V AC, 50-60 Hz, max 0,35 A, Cable length 1,5 m . 8.3. Dimensions Surface area: 66 cm wide, 88 cm deep Seat height:...
  • Page 21 46 | Instructions for use | Reflex Shower-toilet chair...
  • Page 22: Electromagnetic Compatibility

    9. Electromagnetic compatibility Guidance and Manufacturer’s declaration-electromagnetic emissions The Reflex is intended for use in the electromagnetic environment specified below . The customer or the user of Reflex should assure that it is used in the following environment: Emissions Test Compliance Electromagnetic Environment - guidance RF emissions The Reflex uses RF energy only for its internal function . Therefore, its RF Group 1 emissions are very low and are not likely to cause any interference in nearby CISPR11...
  • Page 23 Guidance and Manufacturer’s declaration-electromagnetic immunity The Reflex is intended for use in the electromagnetic environment specified below . The customer or the user of the Reflex should assure that it is used in the following environment: Immunity Test IEC 60601-1-2 Test Compliance Level Electromagnetic Environment - guidance Level Portable and mobile RF communications equipment should be used no closer to any part of the Reflex, including cables than the recommended separation distance calculated from the equation applicable to...
  • Page 24 Recommended separation distance between portable and mobile RF communications equipment and the Reflex The Reflex is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled . The customer or the user of the Reflex can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Reflex as recommended below, according to the maximum output power of the communications equipment .