To Disassemble The Power Supply And Interface Board - Rigol MSO4000 series Service Manual

Digital oscilloscope
Hide thumbs Also See for MSO4000 series:
Table of Contents

Advertisement

RIGOL
To Disassemble the Power Supply and Interface
Board
Power Supply
Interface Board
Figure 1-5 To Disassemble the Power Supply and Interface Board
Part Explanations:
① 4 screws (torx pan head screw #FM3*10) fixing the power supply.
② 3 screws (torx pan head screw assembly #FW3*6) fixing the rear interface board.
Disassemble Steps:
1.
Remove the 2 nuts and washers at the BNC ports on the rear metal cover using
the BNC socket.
2. Remove the British System DB9 screws on both ends of RS232 using the 5mm
hexagon socket.
3. Remove the 4 screw (①) fixing the power supply using the screwdriver (T10).
4. Remove the 3 screws (②) fixing the rear interface board using the screwdriver
(T10).
5. Remove the power supply module and the interface board gently.
1-8
Chapter 1 Disassemble and Assemble
RS232
BNC × 2
MSO4000/DS4000 Service Guide

Advertisement

Table of Contents
loading

Table of Contents