Technical Specifications - ResMed AirFit F30 User Manual

Hide thumbs Also See for AirFit F30:
Table of Contents

Advertisement

WARNING
• Magnets are used in the lower headgear straps and the frame of the mask. Ensure the
headgear and frame is kept at least 2" (50 mm) away from any active medical implant
(eg, pacemaker or defibrillator) to avoid possible effects from localized magnetic
fields. The magnetic field strength is less than 400 mT.
• Avoid connecting flexible PVC products (eg, PVC tubing) directly to any part of the
mask. Flexible PVC contains elements that can be detrimental to the materials of the
mask, and may cause the components to crack or break.
• Regularly clean your mask to maintain the quality of your mask and to prevent the
growth of germs that can adversely affect your health.
• Always follow cleaning instructions. Some cleaning products may damage the mask,
its parts and their function, or leave harmful residual vapours.
CAUTION
• Remove all packaging before using the mask.
• When fitting the mask, do not overtighten the headgear as this may lead to skin
redness or sores around the mask cushion.
• As with all masks, some rebreathing may occur at low pressures.
• Visual criteria for product inspection: If there is any visible deterioration of a mask
component (cracking, crazing, tears etc.), the component should be discarded and
replaced.

Technical specifications

Mask setting options
For AirSense or AirCurve: Select 'Full Face'.
Compatible devices
For a full list of compatible devices for this mask, see the Mask/Device Compatibility List
at ResMed.com/downloads/masks.
Pressure-flow curve:
6
Pressure
Flow
(L/min)
(cm H
O)
2
3
19
10
37
20
54
30
69