7. Ensure that both sample lids of the cartridge are firmly closed. When the cartridge entrance
port on the top of the DiagCORE
barcode facing to the left and the reaction chambers facing down.
There is no need to push the cartridge into the DiagCORE
cartridge entrance port, and the DiagCORE
into the Analytical Module.
8. Upon detecting the cartridge, the DiagCORE
cartridge entrance port and start the test run. No further action from the operator is required to
start the run.
NOTE
The DiagCORE
used and scanned during the test setup. If a cartridge other than the one
scanned is inserted, an error will be generated and the cartridge will be
automatically ejected.
NOTE
Up to this point, it is possible to cancel the test run by pushing the Cancel
button in the bottom right corner of the screen.
NOTE
Depending on the system configuration, the operator may be required
to re-type their user password to start the test run.
®
Analyzer automatically opens, insert the cartridge with the
®
Analyzer will automatically move the cartridge
®
Analyzer will automatically close the lid of the
®
Analyzer will not accept a cartridge other than the one
®
DiagCORE Respiratory Panel 2 Instructions for Use
Analyzer. Position it correctly into the