10. After the test run is completed, the Eject screen will appear and the Module Status Bar will show
the test result as one of the following options:
TEST COMPLETED
TEST FAILED
TEST CANCELED
NOTE
11. Press
with all national, state, and local health and safety regulations and laws.
The cartridge should be removed when the cartridge entrance port opens and ejects the
cartridge. If the cartridge is not removed after 30 seconds, it will be automatically moved back
into the Analyzer, and the lid of the cartridge entrance port will close. If this occurs, press Eject
to open the lid of the cartridge entrance port again and then remove the cartridge.
NOTE
12. After the cartridge has been ejected, the Results Summary screen will appear. Please refer to
section Interpretation of Results for further details. To run another test, press Run Test.
NOTE
24
DiagCORE Respiratory Panel 2 Instructions for Use
The test was completed successfully.
An error occurred during the test.
The user canceled the test.
If the test fails, refer to the DiagCORE Analyzer User Manual Troubleshooting
section for possible reasons and instructions on how to proceed.
Eject to remove the cartridge and dispose of it as biohazardous waste in accordance
Used cartridges must be discarded.
It is not possible to re-use cartridges for tests for which the execution was
started but then subsequently canceled by the operator, or for which an
error was detected.
For further information on the use of the instrument, refer to the DiagCORE
Analyzer User Manual.