2. Nucleic acids are purified from the lysed sample via binding to a silica membrane in the
purification chamber of the DiagCORE
alcohol.
3. The purified nucleic acids are eluted from the membrane in the purification chamber and are
mixed with the lyophilized PCR chemistry in the dried-chemistry chamber of the DiagCORE
cartridge.
4. The mixture of sample and PCR reagents is dispensed into the DiagCORE
chambers, which contain lyophilized, assay-specific primers and probes.
5. The DiagCORE
time PCR and performs real-time fluorescence measurements to generate amplification curves.
6. The DiagCORE
delivers a test report.
14
DiagCORE Respiratory Panel 2 Instructions for Use
Analyzer creates the optimal temperature profiles to carry out effective RT-real
®
Analyzer Software interprets the resulting data and process controls and
®
cartridge in the presence of chaotropic salts and
®
®
cartridge PCR
®