Using the Fluorescence Detector

Instrument Configuration

Instrument Configuration

Instrument Configuration
1 Set the switches of the Configuration switch at the rear of the module:
2 Enter the setup information (MAC / IP address and/or Instrument Name).
58
a All switches DOWN: module uses the default IP address 192.168.254.11.
b Switch 4 UP and others DOWN: module uses DHCP.
c Switch 5 UP and others DOWN: modules uses STORED address
a Agilent OpenLab ChemStation (Configure Instrument):
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual