Bin Display> (Not Available For Lcr-1010) - UNI-T LCR-1000 series Operation Manual

Benchtop lcr meter
Table of Contents

Advertisement

Monitored Voltage and Current
Vm and Im are monitored voltage and current of the DUT respectively. When Vm or Im is selected, it can be
set ON or OFF.
FUNC : Cp-D
FREQ : 1 kHz
LEVEL : 1.00V
/ Cp : -0.0015p
D : 0.3697
Vm:OFF
Use softkeys to select
2-2. <BIN DISPLAY> (not available for LCR-1010)
Press [BIN DISP] key in the pay of <MEAS DISP>, the <BIN DISPLAY> page will be displayed on screen as
shown below:
COMP : OFF
NOMINAL: 0.0000pF
BIN
LOW [ ]
1
2
3
2nd
AUX :
OFF
Cp: -0.0009p F
D : 0.0033
Use softkeys to select
COMP: Set comparator ON or OFF
AUX: Set auxiliary ON or OFF
COUNT: Set counting ON or OFF
NOMINAL: Input nominal value via numeric keys or softkeys. It is reference value of main parameter range.
Deviation is for main parameters only. There are 3 operations to Deviation: OFF, ABS or %.
 ABS (Absolute Deviation): The absolute deviation is the currently measured value minus preset nominal
value. Its calculation formula is:
X is the currently measured value. Y is preset nominal value.
 % (percentage deviation): The percentage deviation is the percentage of the difference between the
currently measured value and the preset nominal value. Its calculation formula is:
Y×100[%]
X is the currently measured value. Y is preset nominal value.
< MEAS DISP >
RANGE : AUTO
SPEED : SLOW
COMP : OFF
Im:OFF
< BIN DISP >
AUX : OFF
COUNT : OFF
MODE : ABS
HIGH [ ]
COUNT
OUT
:
BIN:
ABS = X – Y.
ON
OFF
F
MEAS
SETUP
BIN
DISP
QUICK
CLEAR
% = (X – Y)/