Monitored Voltage and Current
Vm and Im are monitored voltage and current of the DUT respectively. When Vm or Im is selected, it can be
set ON or OFF.
FUNC : Cp-D
FREQ : 1 kHz
LEVEL : 1.00V
/ Cp : -0.0015p
D : 0.3697
Vm:OFF
Use softkeys to select
2-2. <BIN DISPLAY> (not available for LCR-1010)
Press [BIN DISP] key in the pay of <MEAS DISP>, the <BIN DISPLAY> page will be displayed on screen as
shown below:
COMP : OFF
NOMINAL: 0.0000pF
BIN
LOW [ ]
1
2
3
2nd
AUX :
OFF
Cp: -0.0009p F
\
D : 0.0033
Use softkeys to select
COMP: Set comparator ON or OFF
AUX: Set auxiliary ON or OFF
COUNT: Set counting ON or OFF
NOMINAL: Input nominal value via numeric keys or softkeys. It is reference value of main parameter range.
Deviation is for main parameters only. There are 3 operations to Deviation: OFF, ABS or %.
ABS (Absolute Deviation): The absolute deviation is the currently measured value minus preset nominal
value. Its calculation formula is:
X is the currently measured value. Y is preset nominal value.
% (percentage deviation): The percentage deviation is the percentage of the difference between the
currently measured value and the preset nominal value. Its calculation formula is:
Y×100[%]
X is the currently measured value. Y is preset nominal value.
< MEAS DISP >
RANGE : AUTO
SPEED : SLOW
COMP : OFF
Im:OFF
< BIN DISP >
AUX : OFF
COUNT : OFF
MODE : ABS
HIGH [ ]
COUNT
OUT
:
BIN:
⊿
ABS = X – Y.
ON
OFF
F
MEAS
SETUP
BIN
DISP
QUICK
CLEAR
⊿
% = (X – Y)/