Comm Setup; About System - UNI-T LCR-1000 series Operation Manual

Benchtop lcr meter
Table of Contents

Advertisement

ToneSour
Set tone source as MASTER, EAR PHONE or ALL.
PassTone
Set pass tone OFF, or set pass tone as LONG tune, SHORT tune, TWO SHORT tunes.
FailTone
Set fail tone OFF, or set fail tone as LONG tune, SHORT tune, TWO SHORT tunes.
ParaSave
Set parameter save mode as AUTO SAVE, AUTO LOAD, or NO SAVE.
PassWord
In this page, press corresponding softkeys to choose different functions:
1) Set password OFF
2) Lock system: when setting password to lock system, all files are protected and
3) Lock file: it means protecting files.
4) Modify password: the default password is 0010 for LCR-1010, 0030 for LCR-1030 and
5) Save to USB flash disk: The password can be save to USB flash disk. The password
2-5. <COMM SETUP>
Press [COMM SETUP] softkey, <COMM SETUP> page will be displayed on screen as shown below:
Bus Mode
BaudRate
Data Bit
Stop Bit
Parity
Tx Term
GpibAddr
CMD Type
Use softkeys to select
Bus Mode
Set bus mode as RS232C, GPIB, USBTMC or USBCDC.
BaudRate
Use "+" or "-" to choose baud rate from 4800, 9600, 19200, 38400, 115200.
Baud rate is only for bus mode RS232C.
Data Bit
Set data bit as 6, 7 or 8.
Stop Bit
Set stop bit as 1 or 2.
Parity
Set parity as NONE, ODD, or EVEN.
Tx Term
Set end of text term as LF, CR or LFCR.
In response to ASCII code, LF is 0x0A and CR is 0x0D.
This is only end of text term when the LCR meter is returning data to PC.
GpibAddr
Set GPIB address between 0 to 31. Input numbers via numeric keys or softkeys.
This is only for bus mode GPIB.
CMD Type
Please refer details to Programming Manual.
2-6. <ABOUT SYSTEM>
Press [ABOUT SYSTEM] softkey, the <ABOUT> page will be displayed on screen as shown below:
password is required during power on
0200 for LCR-1200.
file name is "0010.STA" for LCR-1010, "0030.STA" for LCR-1030, "0200.STA" for LCR-
1200. Plug USB flash disk into USB HOST, when a password is required during power
on, the LCR meter automatically detects the password file and read it to get password.
< COMM SETUP >
:
RS232C
:
9600
:
8
:
1
:
None
:
LF
:
08
:
SCPI
SYSTEM
SETUP
COMM
SETUP
ABOUT
SYSTEM

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lcr-1030Lcr-1010Lcr-1200

Table of Contents