50MHz,which is ideal for big power testing,development and maintain. The UT-P32 is designed to operate with the 1MΩ impedance oscilloscopes. When combine with the 50Ω load , the attenuation will be 2 times.
Page 8
≤3000V for x1000,(about 1000V RMS and 1500V DC) ≤300V for x100,(about 100V RMS and 150V DC ) (5) Permitted max input voltage: Max differential voltage: 3000V(DC+AC PEAK to PEAK) Max voltage between each input terminal and ground: 5KV RMS (6)Input Impedance: Differential: 8MΩ...
(11)Power Supply:Only External 6V DC power supply. (12)Consumption: 150mA max (0.9 Watt) 3.Operatingenvironmental conditions Reference Storage +20℃…+30℃ 0℃….+50℃ -30℃….+70℃ Temperature Relative Humidity ≤70%RH 10%…85%RH 10%…90%RH (1) Dimensions and weight:69x26x165mm; 500g (2) Electrical safety to IEC 1010-1 Dual insulation Installation category Ⅲ Degree of Pollution 2 Related voltage or max line-earth:5KV RMS CE:EN50081-1 and 50082-1...
․ Select the attenuation ratio* and the vertical deviation of the oscilloscope in accordance with the conversion table below. ․ NB: The POWER light must come on. Attenuation ratio X1000 X100 Voltage Input Range 3000V 300V (DC+AC Peak to PEAK) [N.B] The real vertical deviation in V/div is equal to the attenuation factor multiplied by the range of vertical...
▉ Cleaning This probe does not require any particular cleaning.If necessary,clean the case with a cloth slightly moistened with the soapy water. ▉ Warranty Unless notified to the country,our instruments are guaranteed against any manufacturing defect or material defect. They do not bear the specification known as the safety specification.