Meas Setup; Systems Setup - UNI-T LCR-1000 series Operation Manual

Benchtop lcr meter
Table of Contents

Advertisement

2-3. <MEAS SETUP>
Press [SETUP] key, <MEAS SETUP> page will be displayed on screen as shown below:
Use softkeys to select
ResSource
Set internal resistance source at 30Ω or 100Ω
AVERAGE
Set average number of times between numbers 1 to 255. Input numbers by numeric keys
or soft key "+" and "-".
RangeDwe I I
Set range delay time between 0ms to 6000ms. Input numbers by numeric keys or soft key
"+" and "-".
TrigSource
Set trigger source as INT or EXT.
TrigDelay
Set trigger display time between 0ms to 6000ms. Input numbers by numeric keys or soft
key "+" and "-".
TrigEdge
Set trigger edge as RISING or FALLING.
Handler
Set comparator to CLEAR, HOLD, PULSE.
PulseWidth
Set pulse width between 1ms to 9999ms.
2-4. <SYSTEMS SETUP>
Press [SYSTEM] key, <SYSTEMS SETUP> page will be displayed on screen as shown below:
Theme
Language
Key Tone
ToneSour
PassTone
FailTone
ParaSave
PassWord
Use softkeys to select
Theme
Set display theme to BLUE, BLACK or GREEN.
Language
Set language as English or Chinese
Key Tone
Turn on or off key tone.
< MEAS SETUP >
ResSource
:
30Ω
AVERAGE
:
1
RangeDwe I I
:
0
TrigSource
:
INT
TrigDelay
:
0
TrigEdge
:
RISING
Handler
:
CLEAR
PulseWidth
:
1
< SYSTEM SETUP >
TRAD-BLUE
: 英语
:
ON
:
MASTER
:
OFF
:
TWO SHORT
:
AUTO SAVE
:
OFF
MEAS
SETUP
ms
ms
ms
SYSTEM
SETUP
COMM
SETUP
ABOUT
SYSTEM
SYSTEM
DEBUG

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lcr-1030Lcr-1010Lcr-1200

Table of Contents