Specifications; Measurement Range; Accuracy - UNI-T LCR-1000 series Operation Manual

Benchtop lcr meter
Table of Contents

Advertisement

3. SPECIFICATIONS

3-1. Measurement Range

With 100Ω source resistance, there are five ranges: 31.6Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ.
With 30Ω source resistance, there are six ranges: 10Ω, 30Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ.
The effective measurement range is listed as below.
Table 3-1 Effective measurement range
with 100Ω source resistance
No.
Resistance
0
100kΩ
1
10kΩ
2
1kΩ
3
100Ω
4
30Ω

3-2. Accuracy

3-2-1. Accuracy of │Z│, L, C,R, X
The accuracy (A
) of │Z│,L,C,R,X is shown as below:
e
A
= ±[A+(K
+K
+K
)×100+K
e
a
b
f
A: Basic measurement accuracy (Refer to Fig.3-1)
:Impedance scaling factor (Refer to table 3-4), impedance <500Ω
K
a
:Impedance scaling factor (Refer to table 3-4), impedance >500Ω
K
b
:Temperature factor (Refer to table 3-5)
K
c
:Calibration interpolation factor (Refer to table 4-6)
K
f
:Cable length factor (Refer to table 4-7)
K
L
Note: Choose only Ka or Kb, depending on resistance value. For others, input zero.
Condition for L, C, X accuracy: D
Condition fo R accuracy: Q
When D
≥0.1, for L, C and X, its accuracy factor A
x
When Q
≥0.1, for R, its accuracy factor A
x
Effective mea.range
100kΩ-100MΩ
10kΩ-100kΩ
1kΩ-10kΩ
50Ω-1kΩ
0Ω-50Ω
]×K
[%]
L
c
(D measured value)≤0.1
x
(Q measured value)≤0.1
x
shall be multiplied by
e
Table 3-2 Effective measurement range
No.
Resistance
0
1
2
3
4
5
shall be multiplied by
e
2
1
Q
x
with 30Ω source resistance
Effective mea.range
100 kΩ
100kΩ-100MΩ
10 kΩ
10kΩ-100kΩ
1 kΩ
1kΩ-10kΩ
100Ω
100Ω-1kΩ
30Ω
15Ω-100Ω
10Ω
0Ω-15Ω
2
1
D
x