Section 3: Hardware Error Messages

"Z0 Sensor at Wrong State"

Scenario
Screen Carrier is stuck while the screen is being loaded. As a result the Screen Carrier doesn't reach the Z0 Sensor
(J505) so that it doesn't change its position from "1" to "0".
Hardware and Software Response
[1] Screen unload is halted.
[2] Cassette is still locked in the Reader.
[3] CR Reader operating status changes to "State stalled".
Fix
User Response
[1] Click [OK]; the CR Reader is halted. Status remains "state stalled".
[2] Manually release the cassette using the Release knob under the cassette tray, and remove the cassette.
[3] Reset the CR Reader by turning it OFF and ON again.
[4] Insert the cassette again; does the error repeat?
Yes
Contact Service
Service Response
[1] Remove the CR Reader cover.
[2] Is the Screen Carrier stuck?
a. Manually move the Screen Carrier to its homing position.
b. Turn the CR Reader ON.
c. Insert a cassette and see if one of the sensors is interfering
with the Screen Carrier movement.
Yes
Adjust the sensor position. Check that the Rotational motor
AT000115-02
No
Continue using the CR Reader
Yes
No
cable in the middle of the drum isn't
interfering with the Linear Slide
movement.

Hardware Error Messages

No
Check/replace the Z0 J505 Sensor
according to
"Sensors Tests" on Page 53