Image Artifacts; White Lines Across Image; White Lines Across Image Artifact Examples - Kodak Point-of-Care CR 120 System Troubleshooting Manual

Software version 2.5
Hide thumbs Also See for Point-of-Care CR 120 System:
Table of Contents

Advertisement

Image Artifacts

Section 6: Image Artifacts

White lines across image

White Lines Across Image Artifact Examples

Cause of Artifact
The Linear Slide gets stuck in the middle of a scanning procedure.
There might be a mechanical noise while scanning.
Note
When the Kodak Image Processing option is functioning this artifact is more severe.
Fix
Tighten the tension of the Slide Screw nut. This will also reduce the mechanical noise that accompanies scanning.
Required Tool:
• 12 mm wrench
AT000115-02