Plate Didn't Reach W0 Sensor - Kodak Point-of-Care CR 120 System Troubleshooting Manual

Software version 2.5
Hide thumbs Also See for Point-of-Care CR 120 System:
Table of Contents

Advertisement

"Plate Didn't Reach W0 Sensor"

Scenario
The User inserts a cassette and one of the following occurs:
• The CR Reader fails to load the screen to the drum.
• The screen is loaded to the drum but the CR Reader reports that the W0 sensor (J511) didn't change its position
when the screen was loaded.
• The error message will appear within 5 seconds if one of the above problems occurs.
Hardware and Software Response
[1] Ejects the cassette.
Fix
User Response
[1] Check if the screen is in the cassette.
a. Insert the cassette into the CR Reader.
b. Go to Step 2
[2] Does the error message repeat?
Insert another cassette. Does the error message repeat?
Yes
Contact Service
Service Response
[1] Activate the Roller Motor by means of the Diagnostic screen. Is it turning?
a. Open the back service panel.
b. Check/replace fuse F4 -0.8 A
c. Go to Step 2.
[2] Install Service Pack according to the software version. Check the right Service Bulletin:
• For V2.1.2/V2.2.1 SB000046 - K-QC PoC 120/140
• For V2.4.2
• For V2.5
AT000115-02
Yes
Yes
This is a cassette problem, contact
Service to replace cassette.
Yes
SB000048 - K-QC PoC 120/140/260 SP 3
SB000053 - K-QC PoC 120/140/260 SP 2
No
a. Insert the screen into the cassette. Pay
attention to insert the screen right side up.
b. Insert the cassette into the CR Reader.
c. Go to Step 2
Continue working as usual.
No
Go to Step 2.
SP 2
Hardware Error Messages
No
No