For Pool and Spa Salt Systems
Salt/Chloride (as NaCl) Test Procedure
TR7
TURN METER ON
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: TR7
2
Press and re-press the MENU button until the display shows the parameter TR7.
PREPARE THE CELL
3
Using deionized or distilled water, rinse the CELL at least 3 times to minimize potential cross
ADD THE SAMPLE
4
Use the small pipet provided to add 0.2ml of the sample to the water in the cell. This is a
1 to 20 dilution of the pool water that is required to perform this test. NOTE: If you are using the
more accurate dilution procedure with the Dilution Kit (487200) or Mini Dilution Kit II (487202),
follow the procedure as provided with those kits.
ZERO METER*
5
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Meter is ready for testing.
DIP STRIP
6
Dip the eXact
press READ. This starts the 20 SECOND countdown timer. During this time move the strip
in a gentle back and forth motion. Remove and discard the strip after "1" on the display
disappears*. The cursor will move across the display while the meter prepares to measure
the sample. Record result displayed (this result is automatically stored in TR7).
USE TABLE
7
Find the "TR7" result in the table below to determine the Sodium Chloride concentration in ppm
(parts per million). (Example: a "TR7" result of 65.3 (use only the 65 for the chart) equals a Sodium
Chloride value of 1400 ppm). Record result. After testing is completed, rinse cell immediately.
Sodium Chloride results require the table below. Follow eXact
using eXact
Strip Micro CH, Part No. 481657-II. NOTE: To convert the NaCl value to Chloride (Cl
®
multiply the value from the chart below by 0.6. (Example: 1380ppm NaCl = 828ppm Cl
eXact
%T
9
90
0
40
80
400
440
70
800
840
60
1240
1280
50
1640
1680
40
2040
2060
2400
2460
30
2820
2860
20
3280
3340
10
3940
4040
0
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
Strip Micro CH (as NaCl), Part No. 481657-II into the CELL and immediately
®
Chloride
Strip Micro CH, Part No. 481657-II - for 4mL Samples
®
8
7
6
80
120
480
520
880
960
1320
1360
1720
1760
2100
2140
2500
2540
2900
2940
3400
3460
4200
4280
Table
(as NaCl)
Micro 7+ Chloride (as NaCl) Test Procedure (above)
®
5
4
160
200
560
600
1000
1040
1400
1440
1800
1840
2180
2200
2580
2620
2980
3040
3520
3580
4440
4660
14
3
2
240
280
320
640
680
720
1080
1120
1160
1480
1520
1560
1880
1920
1960
2240
2280
2320
2660
2700
2740
3080
3140
3200
3640
3700
3760
5000
5840
7000
Rev. 041311-BT
CH
),
–
)
–
1
0
360
760
1200
1600
2000
2360
2780
3240
3840
>7000