Compliance West HT-3000 Instruction Manual

Dielectric withstand tester. 0-2000 volts ac output and 0-2800 volts dc output

Advertisement

Quick Links

HT-3000
Dielectric Withstand Tester
0-2000 Volts AC Output
0-2800 Volts DC Output
Instruction Manual
COMPLIANCE
WEST
USA

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HT-3000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Compliance West HT-3000

  • Page 1 HT-3000 Dielectric Withstand Tester 0-2000 Volts AC Output 0-2800 Volts DC Output Instruction Manual COMPLIANCE WEST...
  • Page 2 To fully appreciate all the features of your new meter, we suggest that you take a few moments to review this manual. Compliance West USA stands by your instrument with a full one- year warranty and a loaner instrument policy. If the need arises, please don't hesitate to call on us.
  • Page 3: Table Of Contents

    Table of Contents An Introduction to Dielectric Withstand Testing with the HT-3000............1 Safety Precautions........................1 Test Personnel........................1 Testing Area........................1 Safety Techniques......................2 Using the HT-3000 Dielectric Withstand Tester................2 Ground Check........................2 Ground Check Failures..................2 Defeating the Ground Check ................3 Leakage Test........................3 Sporadic Leakage Current Failures ..............3 Chronic Leakage Current Failures..............3...
  • Page 4 Maintenance and Calibration........................20 Service Information........................20 General Maintenance ........................20 Interior Access .......................20 Calibration Access...................20 Cleaning.........................21 Performance Test .........................21 Operation/Function Test ....................21 Calibration Procedure ........................22 Meter Calibration Adjustment ..................22 Continuity Calibration Adjustment................23...
  • Page 6: An Introduction To Dielectric Withstand Testing With The Ht-3000

    Currents of as little as 5 mA at 120 volts can cause death, and the HT-3000 is capable of generating 20 mA AC at up to 2000 volts, and 5 mA DC at up to 2800 volts. The HT-3000 has been designed to minimize exposure to high voltages.
  • Page 7: Safety Techniques

    The equipment under test should be connected to the HT-3000 and then left alone by the operator. After the operator is clear of the Tester and the equipment under test, he should press the RESET Button, then the TEST Button, with his right hand.
  • Page 8: Defeating The Ground Check

    If the Test Return Lead is not properly connected, a dielectric withstand failure may energize exposed dead metal. Additionally, the HT-3000 may not recognize the failure. The test will continue for its normal length of time, and the HT-3000 may show a "PASS".
  • Page 9: High Voltage Dielectric Withstand Test

    If the the HT-3000 is being used to conduct an AC test and the leakage current adjustment is set to 20 mA and almost all examples of the model being tested fail, the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used. This is due to the AC voltage charging all capacitors connected between primary voltage and ground.
  • Page 10: Introduction And Specifications

    A 120kohm resistor is supplied on the rear Introduction panel for leakage current checks. This manual contains complete operating, maintenance and calibration instructions for the The HT-3000 meets all safety agency criteria for Compliance West Model HT-3000 automatic production line dielectric withstand Dielectric Withstand Tester.
  • Page 11 ENVIRONMENTAL Operating Temperature 15-40°C Relative Humidity Range 0-90% non-condensing GENERAL Input power requirements Model HT-3000: 114-127 volts, 50/60 Hz Model HT-3000L: 98-120 volts, 50/60 Hz Weight 10½ lbs. SAFETY AGENCY TOPICS Transformer Output < 500VA Visual Indication of Voltage Output...
  • Page 12: Operation

    The front panel • A Power Cord to provide power to the features of the HT-3000 are shown in Figure 2 HT-3000. and described in Table 2. The rear panel features of the HT-3000 are shown in Figure 3 •...
  • Page 13 Figure 2. Controls, Indicators, Connectors - Model HT-3000 Front Panel...
  • Page 14 RESET Button / Red Indicator When lit, indicates that the HT-3000 is unarmed. This button must be pushed before the TEST Button is functional. When the RESET Button is pressed, the red RESET indicator goes out and the yellow TEST indicator is lit.
  • Page 15 Figure 3-2. Controls, Indicators, Connectors - Model HT-3000 Rear Panel...
  • Page 16 Use supplied cordset to connect Tester to appropriate source of supply. Replace line fuse. Turn Tester Power Switch ON/OFF. Fuse replacement warning / Specifies replacement fuse and supply voltage used. Rating of supply Table 3 Control, Indicators, Connectors - Model HT-3000 Rear Panel...
  • Page 17: Initial Checkout Procedure

    The tester should conduct a test sequence, supply using the included 18 AWG power terminated with a buzzer. At test supply cord. Turn the HT-3000 on. termination, the Ground OK, Ground Turn the Ground Check switch on the Open, Excess Leakage, and red RESET rear panel to Off.
  • Page 18: Setting Up The Ht-3000 For Production Line Testing

    AC Leakage Current can be set to levels Passage of these tests indicates that the of up to 20 mA. DC Leakage Current can be set HT-3000 is functioning properly and that it is to levels of up to 5 mA. safe to use.
  • Page 19: Adjustment Of The High Voltage Ramp Time

    Remove the test leads from the front rear panel Off. panel. Check the position of the Ground Connect the HT-3000 to a correctly rated Check switch. After this procedure, the source of supply and turn the power high voltage level is to be reset using the switch to the I or ON position.
  • Page 20: Adjustment Of The High Voltage Test Duration

    (usually 1 sec. vs. 1 min.) if the voltage is switch to the I or ON position. Push the increased by 20%. The HT-3000 is factory set RESET button. The yellow TEST for 1200 volts AC, a voltage level used for the...
  • Page 21: Operating Techniques

    I or ON position. Push the safety and to ensure that the HT-3000 properly RESET button. The yellow TEST reports all failures. indicator should light, indicating that the The small size of the Ground Check HT-3000 is ready to test.
  • Page 22: Test Verification Procedure When Not Using The Htt-1 Function Checker

    Duration procedure. checker: breakdown insulation system between primary Connect the HT-3000 to a correctly rated voltage and ground is detected, the source of supply and turn the power Hipot Fail LED will light, the buzzer switch to the I or ON position. Push the...
  • Page 23 Leakage Current Fail light terminating the test. Consider lengthening the ramp time; see Adjustment of the High Voltage Ramp Time.
  • Page 24: Technical Assistance

    Technical Assistance For Technical Assistance Phone: (800) 748-6224 Technical Assistance is available from Compliance West USA between the hours of 8:30 AM and 4:00 PM Pacific Time. Compliance West USA 2120 Jimmy Durante Blvd., Suite 124 Del Mar, CA., 92014...
  • Page 25: Maintenance And Calibration

    Calibration Access Service Information The HT-3000 is warranted to the original Use the following procedures to gain access to purchaser for a period of 1 year. This warranty the calibration adjustments of your instrument. does not cover problems due to misuse or neglect.
  • Page 26: Cleaning

    Connect the Tester to a proper source of accessible. supply using the included 18 AWG power supply cord. Turn the HT-3000 on. NOTE Turn the Ground Check switch on the With the power cord replaced, the rear panel to Off.
  • Page 27: Calibration Procedure

    Connect the red lead of the DMM to one Passage of these tests indicates that the of the vertical slots of the DEVICE HT-3000 is functioning properly and that it is receptacle, and the black lead of the DMM safe to use.
  • Page 28: Continuity Calibration Adjustment

    Push the RESET button, then the TEST increased to allow the test to continue for a button. longer time. Set the HT-3000 AC/DC switch to AC. The Ground OK or Ground Open LED Push the red RESET button, then the will immediately light. Push the RESET switch.

This manual is also suitable for: