YOKOGAWA WT500 User Manual page 2

Power analyzer pulse output function (/p17 optional)
Hide thumbs Also See for WT500:

Advertisement

2.2
Power-on Settings and Configuration Switches
When the power is turned on, the voltage range is 100 V, the frequency filter is on, the key lock is
on, the synchronization source is the voltage signal, and the current range and wiring are set to
the values according to the following switch settings. All other settings are the same as the default
settings.
Switch setting
No. Setting
1
Wiring
2
3
Current range
4
Scaling
5
Output pulse source
6
element for single-
phase two-wire
(valid for single-
phase two-wire)
7
Pulse output format OFF: Open collector output
8
Reserved
2/5
→ ON
Description
No.1/2
OFF/OFF: Single-phase two-wire
ON/OFF: Single-phase three-wire
OFF/ON: Three-phase three-wire
ON/ON: Prohibited
OFF: 5 A
ON: 500 mA
OFF:
Scaling off
CT ratio 1
ON:
Scaling on
CT ratio 40
No.5/6
OFF/OFF: Element 1
ON/OFF: Element 2
OFF/ON: Prohibited
ON/ON: Prohibited
ON: Voltage output (0 V to 5 V)
No effect on settings
1P2W (single-phase two-wire)
1P3W (single-phase three-wire)
3P3W (three-phase three-wire)
Prohibited
5 A
500 mA
Disable the scaling function (OFF),
CT ratio setting 1
Enable the scaling function (ON),
CT ratio setting 40
The measured value of element 1
The measured value of element 2
Prohibited
Prohibited
Open collector output
Voltage output (0 to 5 V)
IM760201-52E