Appendix VI: Firmware Revision History
Firm-
ware
Rev.
2.0
Jan. 1, 2002
2. 1
Feb. 5, 2002
2.2
Aug. 27, 2002
2.3
Jan. 23, 2003
2.4
Apr. 23, 2003
3.0
Apr. 1, 2004
Page 102
SITRANS LR400 (7ML5421) – INSTRUCTION MANUAL
Date
•
Original software release
•
Corrected HART board reset.
•
Updated many menu items.
•
Improved multiple echo detection when only two
echoes occur.
•
Resolution of TVT increased.
•
Leading edge detection improved.
•
TVT is no longer reset on factory reset.
•
Sensitivity of the display-mounted keys increased.
•
Level damping max value increased.
•
Silo2 default settings changed.
•
Corrected S/N value of over 700 being displayed
when the auto noise order was set to 0
•
First Echo algorithm improvement
•
Sensitivity of the display-mounted keys reduced.
•
Alarm mode relays now latches during alarm.
•
Removed support for the front keypad; added
support for the hand programmer.
•
Added dynamic TVT hover level feature.
•
Added translations to support both English and
German.
•
Moved all the USER1 and USER2 service fields to
the standard DD under new folder called
"Advanced Echo Processing"
•
Added support for new parameters (hover level
and window trigger)
•
Changes to the echo profile graph in SIMATIC
PDM to correspond to other Siemens Milltronics
products.
•
Cleanup/addition of default values and help
messages
•
Added manual TVT shaper command to SIMATIC
PDM.
Changes
7ML19985FH06