Quick Reference Guide
Tissue-Tek® Glas™ g2
Start Up
®
TM
The Tissue-Tek
Glas
g2 is a system that dispenses
mounting medium onto stained histopathological and
cytopathological samples on glass slides as well as
coverslipping these samples using cover glasses.
Preparation before cover slipping
Hardware
Placing cover glasses
1.
Put cover glass holder in place.
2.
Make sure the tab of the cover glass holder fits in the
holder (check on the CG size).
Placing mounting medium
When setting the mounting medium bottle for the
first time, you must fill the discharging channel
with mounting medium.
1.
Place the waste bottle below the dispensing
nozzle.
2.
Fill the priming bottle with mounting medium.
3.
Connect the tip of the introduction tube with
the priming bottle.
4.
Squeeze the priming bottle to feed the
mounting media until it reaches the dispensing
nozzle.
5.
Remove the priming bottle from the
introduction tube.
6.
Set the mounting medium bottle.
7.
Tilt the mounting medium tube towards the
front and lift the tube.
8.
In order to prevent air bubbles, insert the tube
gently into the bottle.
9.
Perform an [initial priming] page 3.3 OM.
10. Press [menu].
11.
Highlight [extended prime] and press [enter].
Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL
Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | sakura@sakura.eu | www.sakura.eu
Cover glasses
Cover glass holder
Placing cover glasses
Placing mounting medium
Holder hole
Tab
Holder hole
Introduction tube
Waste bottle
Priming bottle

Advertisement

loading
Need help?

Need help?

Do you have a question about the Tissue-Tek Glas g2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sakura Tissue-Tek Glas g2

  • Page 1 10. Press [menu]. Highlight [extended prime] and press [enter]. Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | sakura@sakura.eu | www.sakura.eu...
  • Page 2 Quick Reference Guide Tissue-Tek® Glas™ g2 Changing mounting medium When changing the type of mounting medium, the current mounting medium must be flushed out from the discharging channel before introducing the new Introduction tube medium. Place the waste bottle below the dispending nozzle.
  • Page 3 Quick Reference Guide Tissue-Tek® Glas™ g2 Filling anti drying bottle Fill the anti drying bottle with solvent to just below the lid. Note: type of solvent depends on the mounting medium, most likely xylene Place anti-drying bottle waste container in the dispensing tray. Handling the dispensing nozzle Required level When setting the anti-drying bottle, push down...
  • Page 4 Quick Reference Guide Tissue-Tek® Glas™ g2 SPEED Moves cursor inside name field – this field is surrounded by two brackets. VOLUME This key moves the cursor to highlight the desired characters. ENTER Use this key to select and display the highlighted character in the name field.
  • Page 5 Quick Reference Guide Tissue-Tek® Glas™ g2 Coverslipping Operation Cover slipping slides Pre-operational checks Turn on POWER. Remove the loading station lid. Check the solvent volume level in anti-drying bottle and dispensing pump reservoir. Check CG size Clean any possibly spilled mounting medium. Place receiving racks in the unloading area.
  • Page 6: Maintenance

    Quick Reference Guide Tissue-Tek® Glas™ g2 Maintenance (section 5 operating manual) Daily Station cleaning Clean each station and its surrounding area. Brush off each station, remove foreign matters. Using a cloth moistened with solvent for mounting medium, remove mounting Station cleaning medium attached to the stations.
  • Page 7 The coverslipping applicator pressure is too Increase the applicator pressure of the cover weak and needs to be increased. glass during coverslipping. For details, consult the CSC or Sakura dealer. Small air bubbles are Mounting medium did not spread effectively Decrease the coverslipping speed.
  • Page 8 26 mm (width) x 76 mm (length) x 0.9 to 1.2 mm (thickness) Adjust the system according to the slide glass used. For details, consult the CSC or Sakura dealer. Mounting medium has solidified on or in the Remove the solidified mounting medium.
  • Page 9 The cover glass application pressure during Increase the coverglass application pressure. coverslipping is too weak. For details, consult the CSC or Sakura dealer. The starting position of the application of Change the start position for application of mounting medium not correct.
  • Page 10 The application pressure of the coverglass Decrease the coverglass application pressure. becomes dirty with during coverslipping is too strong. For details, consult the CSC or Sakura dealer. mounting medium. The start position for the dispensing of Change the start position of the dispensing of mounting medium is not appropriate.
  • Page 11 Mounting medium at the tip of the Dissolve the solidified mounting medium using dispensing nozzle has solidified. solvent (usually xylene). For details, consult the Customer Service Center or the Sakura distributor. The system suddenly A power outage occurred. Take a remedial action according to “Actions to stopped operating.