PARAMETER SETUP: AUTO READ MODE

The Auto Read option is available from any parameter Setup. Manual is most typical and the
recommended setting. It is also the factory default setting. Auto Read mode setting is effective only in
Single Channel display.
Auto:
When the Auto Read function is set to AUTO, the meter will freeze the measurement as soon as the
STABLE indicator appears. Once locked, the reading will not change until Measure is selected.
Tip: Take caution when using AUTO; if accidentally selected, it may seem like the
electrode or meter is un-responsive when in fact the measurement is locked and
selecting "Measure" screen is the only way to get the current reading to re-appear.
Manual:
When the Auto Read function is set to MANUAL, the STABLE indicator will appear, however the live
measurement will continue to update on the display. The measurement does not freeze. The "Measure"
icon is not active during Manual setting.
Note: For ISE, if the Ion method is Known Addition, Known Subraction, Analate Addition, and Analate
Subtraction then the Auto Read functionality is inactive. The measure button in the measure button in the
Measurement screen will be replaced with Respective Ion method symbol-either; KA, KS, AA or AS.