5
REMOVE STRIP
Remove one eXact
and set in a dry, convenient place. Replace cap.
6
DIP STRIP AND PRESS READ
Press
simultaneously
gently touching the bottom of the cell. Use a gentle
constant back and forth motion (2 strokes per
second) until the timer displays "1". Be careful to not
spill the sample from the CELL. Remove and discard
the strip.
For water temperatures above 95°F/35°C (hot tubs), Press READ
when the timer displays "10", or use the specially calibrated SPA
eXact
EZ photometer.
®
7
CAP CELL AND READ RESULTS
Place the Cell Cover onto the CELL and
result displayed as Total Alkalinity. This result is
automatically stored in the AL(1) menu. After testing
is complete, rinse the sample cell immediately and
clean with brush.
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
Extender Strip (part no. 486665) until a numerical value is displayed;
Extender Strip (part no. 486665) until a numerical value is displayed;
Count how many AL Range Extender strips were used; each strip
Count how many AL Range Extender strips were used; each strip
adds 130 ppm to the fi nal result.
adds 130 ppm to the fi nal result.
FILL, DIP, READ
TOTAL ALKALINITY
®
to initiate a 20 second countdown and
DIP
exactmeters.com
Strip Micro AL (Part No. 486641)
the eXact
strip in the sample,
®
5
TIP
READ
TIP