3.4.2 Other Measurement Mode Cannot be Used
When a measurement mode is set, only the configured mode can be used. Even if you select 2. MEASUREMENT
on the menu screen, a screen that allows you to select a measurement mode will not display. In stead you will
see the screen that allows you to configure the measurement setting for the set measurement mode. In order to
use other measurement mode, you need to deactivate the current selected mode.
[Example] When Flow Rate mode is set;
POWER SW ‒ ON
Air Volume Calculation
Measurement Standby

3.4.3 How to Deactivate Pre-set Measurement Mode

In the same manner as when you configure program setting, display the <PROGRAM SET> screen.
When it is displayed, select 1. OFF.
Display
< PROGRAM SET >
1.
OFF
2.
AVERAGE
3.
INTERVAL
4.
FLOW RATE
< PROGRAM SET >
1.
OFF
2.
AVERAGE
3.
INTERVAL
4.
FLOW RATE
< MENU >
1. MONITOR
2. MEASUREMENT
3. DATA OUTPUT
4. MEMORY CLEAR
5. UTILITY
1.800.561.8187
Even if you select 2. MEASUREMENT, ·················································
<MENU>
1. MONITOR
2. MEASUREMENT
3. DATA OUTPUT
MENU
SET
Key
(1) Select 1. OFF.
△, ▽
SET
Then press SET key.
OFF
(2) 1. OFF will be highlighted.
Then, the menu screen will be displayed automatically.
(3) Select 2. MEASUREMENT to perform a measurement.
△, ▽
SET
www.
<FLOW RATE>
1. S-TIME (S) 01
2. DATA (M)
010
3. POINTS
016
Procedure
information@itm.com
.com
The below screen will not
be displayed.
<MEASUREMENT>
1. AVERAGE
2. INTERVAL
3. FLOW RATE
21