3. Measurement Mode
INT (min):
POINTS:
* For S-TIME(S), DATA(N), MEMORY, PRINT, refer to page 15 describing the average measurement.
(b)
<INT>
T: 00/10
N: 000/010 P:016
16.8 m/s
127 ℃
(6-1)
<INT>
T: 01/10
N: 001/010 P:016
16.8 m/s
128 ℃
(6-2)
<INT>
WAIT
POINTS
12.8 m/s
126 ℃
(6-3)
<INT>
WAIT
POINTS
16.8 m/s
126 ℃
<INT RESULT>
AVE:
19.8 m/s
MAX:
21.6 m/s
MIN : 18.5 m/s
AVE:
126 ℃
MAX:
129 ℃
MIN :
123 ℃
1.800.561.8187
Set how often a measurement to be started.
Configurable sampling interval (min):
* Interval (INT) has to be set longer time than sampling time at each point.
Set how many points you want to perform a measurement?
Configurable number of points: 1 ~ 999 points
ready
M:036
START/STOP
M:036
[Start
Measuring]
[Waiting]
060 min
001/024
001 min
001/024
[Stop
Measuring]
Calculation
Result
Printing
Example
P24
www.
[Measurement Condition]
1 ~ 6, 10, 12, 15, 20, 30, 40, 50, 60, 100, 120, 150, 180 min.
(5)
ready
will be displayed indicating that the instrument is ready
to start a measurement.
Press START/STOP key to start sampling.
☆ To go back to the measurement setting screen (a), press △
key.
(6)
(6-1):
Sampling at one point.
(6-2):
After a sampling at one point is finished,
the instrument will standby for the next
sampling. On the screen, the average
value for one point will be displayed until
the next measurement will start.
(6-3):
One minute is left before the next
measurement starts
One minute later, the screen will
become screen (6-1) and a measurement
will start.
(7) Repeat the above procedure from (6-1) to (6-3) for the
configured number of points.
☆To stop measuring halfway through: START/STOP or ▽ key.
- The data collected until the measurement is halted will be
calculated.
☆To stop measuring (1): △ key.
- It will go back to the screen (b) where the instrument is
ready to start a measurement.
☆To cancel a measurement (2): MENU key.
- It will go back to the menu screen.
Data at each point is an average data. Average, maximum and
minimum of the average data at each point will be displayed.
☆ PRINT:
☆ START/STOP or ▽: ready for measuring (screen b)
☆ MENU:
.com
As for "ready" display,
refer to page 16
describing average
measurement.
WAIT 060 min:
remaining time length
before the next
measurement starts (min)
POINTS 001/024: point
hardcopy of the display screen
menu will be displayed
information@itm.com
17
waiting time
Total set No. of
measurement trials
Current sampling
trial No.