Model Kern Cfs 50K-3 - KERN CCS 6K-6 Operating Manual

Counting balance/counting system
Hide thumbs Also See for CCS 6K-6:
Table of Contents

Advertisement

 Ensure that no objects are on the weighing plate on
display „LoAd 0".
Wait for stability display, then press
 Carefully place fourth adjustment weight into the center of
the weighing plate once more on display „LoAd 4".
Wait for stability display, then press
 Carefully place third adjustment weight into the center
of the weighing plate once more on display „LoAd 3".
Wait for stability display, then press
 Carefully place second adjustment weight into the center
of the weighing plate once more on display „LoAd 2".
Wait for stability display, then press
 Carefully place first adjustment weight into the center of
the weighing plate once more on display „LoAd 1".
Wait for stability display, then press
 Ensure that no objects are on the weighing plate on
display „LoAd 0".
Wait for stability display, then press
 After the adjustment the balance will carry out a self-test.
The balance returns automatically into weighing mode.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message (
), repeat adjustment process.

15.2 Model KERN CFS 50K-3

 Switch-on balance and during the selftest press
CFS/CCS-BA-e-1825
Operation
.
.
.
.
.
.
„Pin"
.
Display
67

Hide quick links:

Advertisement

Table of Contents
loading