19.1 Error messages

Error message
Err 4
Err 5
Err 6
Err 19
FAIL H / FAIL L
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
84
Description
Zeroing range exceeded
due to switching-on balance
or pressing
(normally
4% max)
Keyboard error
Value outside the A/D
changer range
Zero point displaced
Adjustment error
Possible reasons/
solution
• Object on the weighing
plate
• Overload when zeroing
• Inappropriate
adjustment
• Damaged weighing cell
• Damaged electronics
• Unpurposeful
operation of the scale
• Weighing plate not
installed
• Damaged weighing cell
• Damaged electronics
• Remedy:
Adjust / linearize
• Inappropriate
adjustment
CFS/CCS-BA-e-1825