2 Start of Operation

2.1 Transport and Assembly

The unit may be placed on any flat surface. It must not be placed on the floor. Keep a
wall distance of at least 20 cm. The device should neither be placed in front of radiators
nor should it be covered by pillows or blankets while in operation. The device is not
made for outdoor operation.
For maximum usability there is a special device cart and also a special HiToP
available (see chapter 10, Accessories) which answers in design and function to the
HiToP
4 touch. This device cart is perfect for the placement of the unit and various
accessories. Placed on top of the device cart, the HiToP
ready for use. The device cart can easily be moved next to the patient for the treatment.
With the HiToP
bed patients as well as health care professionals benefit from functions
like faster application of patient cables and electrodes, heated surface of the bed and the
possibility to warm up the electrodes.
The HiToP
4 touch corresponds to the regulations DIN EN 60601. It is a device of
protection class I. Within the scope of the Medical Device Directive (MDD) the current
stimulation device belongs to class IIa.
Warning!
The unit is not designed to be operated in places with the inherent risk of
explosions. If it is used in dangerous areas of anesthesia departments, the
possibility of an explosion cannot be excluded.
If the patient and/or the patient cable is directly exposed to a radiator of a
medical device for high frequency heat therapy, damage of the device or
danger to the patient cannot be excluded. As a rule, a clearance distance
of 3 m is sufficient.

2.2 Connection and Switch-On

The HiToP
4 touch is designed to be connected to a supply voltage of 100 to 240 V. It
is not necessary to switch over the voltage – the device adjusts automatically to the right
voltage.
Irrespective of the adjusted supply voltage, the device is appropriate for mains
frequencies of 50 to 60 Hz.
Connect the HiToP
The protective ground must work correctly.
The HiToP
4 touch is switched on by the main switch on the back of the device.
 gbo Medizintechnik AG
4 touch with the mains cable to a socket with protective ground.
4 touch is mobile and always
bed
Version 1.3