4.2

Setting SimulFAM i

If the current form SimulFAM i is selected with its frequencies of 4096 and 16384,
the current intensity must be such that the patient feels a light tingling only. It is the
aim of this setting to emit the maximum output during therapy. The frequencies of
4096 – 32768 Hz are being emitted as quarter tone steps every second. The current
should be comfortable to the patient with all frequencies. There is no stimulation of
the nerve or muscle, the patient feels a light tingling only. In general, the current form
SimulFAM i is applied for whole body treatment however, it may also be used for a
local lateral throughflow.
Note!
!!

4.2.1 Setting of f-sens

If both intensity values at 4096 Hz and 16384 Hz are set and the therapy is started,
the calculated value for f-sens is indicated. This value indicates the relation of the
current value at 16384 Hz relative to the current value at 4096 Hz.
As already mentioned above the patient should feel a tingling during the therapy with
all 72 frequencies running. In the event that the tingling cannot be felt equally, it can
be improved with a change in f-sens. When pressing the button f-sens, the device will
switch into a faster run with a lower amount of frequencies. With the aid of the arrow
key you can set the upper and the lower frequencies equally.
After a short timeout the regular frequency scan will restart. Now one can judge
whether the adjustment improved which means that with all frequencies almost an
identical current feeling is perceived.
4.3

Setting SimulFAM X

If the current form SimulFAM X is selected, the frequencies are run through faster.
The frequency run can be set between 01 Hz up to 200 Hz.
 gbo Medizintechnik AG
 The amplitude values move on the threshold curve determined.
 Therapy time may also be modified during treatment.
 With the frequency button the frequency flow may be stopped
and released upon new activation.
 The frequency set may be changed with the bar beneath the
graphics.
 The intensity of SimulFAM i may also be changed during
therapy.