24

5.3 Electrode positioning

The position of the patient is very important. The therapy shall always be carried out
while the patient is in a comfortable and relaxed position. The joints should be
bended in a mid-position to allow both the flexor as well as the extensor muscles to
be stimulated.
In general, the preparation and the application of the electrodes are carried out as
follows:
1. Moisten the conductive rubber electrodes with contact spray to avoid current
sensations on the skin. Apply two puffs of the spray to the electrodes and
spread the liquid over the electrodes' surface.
2. Apply the electrodes to the patient.
3. Apply the elastic strap with Velcro in such a way that the electrode covers the
skin all-over.
4. Connect the electrodes to the patient cable.
5. Connect the patient cable to the distribution cable.
6. Plug in the patient plug of the distribution cable to the respective channel
socket of the unit.
Warning!
 Do not apply the electrodes on skin injuries. Even minor abrasions
can cause a burning sensation to the patient. Consequently, the
intensity of the current may be judged erroneously by the patient. If
an electrode must be placed in spite of it all, it is advised to apply
zinc ointment or white petrolatum to cover the affected parts of the
skin.
 The electrodes must cover the skin all-over in order to avoid
excessive local current densities. Otherwise, flush or even burns
may show.
 During the therapy the patient must take off jewelry and eyeglasses.
Note!
!!
 A current density of over 2 mA/cm
increased attention of the operator.
 gbo Medizintechnik AG
2
on all electrode surfaces requires
Version 1.3