Guidance and manufacturer's declaration — electromagnetic immunity
The equipment is intended for the use in the electromagnetic environment specified below. The customer or the user
of the equipment should ensure the compliance with this stipulation.
Immunity test
conducted RF,
IEC 61000-4-6
radiated RF,
IEC 61000-4-3
Recommended separation distances to portable and mobile RF communication
equipment
The equipment is intended to be operated in an electromagnetic environment where radiated RF interference is
controlled. The user can help to avoid interferences by meeting the minimum separation distances between portable
and mobile RF communication equipment (transmitters) according to the maximum output power of the
communication equipment.
Rated power of the
transmitter (W)
0.01
0.1
1
10
100
 gbo Medizintechnik AG
IEC 60601- test level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Separation distance according to the transmission frequency (m)
150 kHz to 80 MHz
d=1.2P
0.12
0.38
1.2
3.8
12
Compliance level
Portable and mobile RF communications
equipment should be used no closer to any
part of the equipment, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
3 V
d=1.2P
eff
3 V/m
d=1.2P
for 80 MHz to 800 MHz
d=2.3P
for 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
80 MHz to 800 MHz
d=1.2P
0.12
0.38
1.2
3.8
12
Electromagnetic environment –
guidance
800 MHz to 2.5 GHz
d=2.3P
0.23
0.73
2.3
7.3
23
39