5.4 The Single-Traverse Method - UCF

Note: The instructions in this section can also be used for a multiple-traverse method. However, you must use an
ODD number of traverses. Consult the factory for details.
Note: You will need both the short block assembly and the long block for a single-traverse installation. You will also
need a marker or scribe to mark the transducer locations on the pipe
To install a single-traverse UCF, complete the following steps:
1. Obtain the transducer spacing dimension, as described in your flowmeter User's Manual.
2. Be sure the location you have chosen for the installation has a straight pipe and undisturbed flow for at least 10
pipe diameters upstream and 5 pipe diameters downstream from the flow measurement point.
3. Prepare the pipe where you intend to place the clamping fixture by making sure it is clean and free of loose
material. Sanding, though usually not required, may be necessary to remove any high spots on the pipe. When
sanding, be careful to preserve the original curvature of the pipe.
4. Find the top of the pipe and use a level to draw a line parallel to the center line of the pipe.
C-RS Ultrasonic Flow Transducer Installation Guide
Line
Top of Pipe
Side View
11