Resetting Alarms And Faults - Honeywell Sensepoint XRL Instruction Manual

Fixed gas detector
Hide thumbs Also See for Sensepoint XRL:
Table of Contents

Advertisement

Maintenance

4.7 Resetting Alarms and Faults

38
The default setting of each alarm is non-latching type. If you want to
have a latching type alarm, change the alarm mode using the
Sensepoint mobile app.
1. Pair with the target detector and tap Settings.
2. Tap the edit button
3. Scroll down to Alarm latching and choose Latching.
4. Scroll down to the bottom of the page and tap Apply Settings to
save the change.
When there are latched gas alarm or faults, reset them using either of
these two methods:
• Use the mobile app to pair with the detector, and tap Maintenance >
Reset Alarm.
• Alternatively, press the UP or DOWN button shortly.
This will reset latched alarms and faults, and initiate a self diagnostics
to make sure there is no warning or fault in the detector.
at the upper right side of the screen.