Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations /
Complies
Flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Immunity Test
Electrostatic Discharge (ESD)
IEC61000-4-2
Electrical fast transient/ burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to the application of the test level
T

Appendix A: EMC Information

The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network.
IEC60601 test level
±6kV contact
±6kV contact
±8kV air
±8kV air
±2kV for power supply
±2kV for power supply
line
line
±1kV for input/out line
±1kV for input/out line
±1kV for differential
±1kV for differential
mode
mode
±2kV for common
±2kV for common
mode
mode
<5 % U
(>95 % dip in
<5 % U
T
U
)for 0,5 cycle
U
T
40 % U
(60 % dip in
40 % U
T
U
)for 5 cycles
U
T
70 % U
(30 % dip in
70 % U
T
U
)for 25 cycles
U
T
<5 % U
(>95 % dip in
<5 % U
T
U
)for 5 sec
U
T
3 A/m
3 A/m
Electromagnetic Environment-Guidance
Compliance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of atypical commercial or hospital
environment
Mains power quality should be that
of atypical commercial or hospital
environment.
(>95 % dip in
Mains power quality should be that
T
) for 0,5 cycle
of atypical commercial or hospital
T
(60 % dip in
environment. If the user of this
T
) for 5 cycles
device requires continued operation
T
(30 % dip in
during power mains interruptions, it
T
) for 25 cycles
is recommended that the device be
T
(>95 % dip in
powered from an uninterruptible
T
) for 5 sec
power supply or a battery.
T
Power frequency magnetic fields
should be at levels characteristic of
atypical
commercial
environment.
18
Electromagnetic
Environment-Guidance
location
in
a
typical
or
hospital