Appendix A: Emc Information - Apex Digital XT Auto User Manual

Table of Contents

Advertisement

Available languages

Available languages

XT Auto/ English

Appendix A: EMC Information

Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Warning:
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the CPAP, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Immunity Test Levels
Basic EMC
Professional
healthcare
standard
facility environment
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function.
Group1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
Class B
The device is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the
public low-voltage power supply network
Complies
HOME
HEALTHCARE
ENVIRONMENT
Compliance
Electromagnetic
Levels
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
±8kV contact
floors are covered with
±15kV air
synthetic material, the relative
humidity should be at least 30
%.
±2kV for power
supply line
Mains power quality should be
±1kV for
that of atypical commercial or
hospital environment
input/output
line
19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

9s-005720

Table of Contents